Measurement of the sensitive profile in a solid state silicon detector, irradiated by X-rays
Mid Sweden University, Faculty of Science, Technology and Media, Department of Electronics Design. (Radiation Sensor Systems)
RTI Electronics AB, Lund University IKVM.
Mid Sweden University, Faculty of Science, Technology and Media, Department of Electronics Design. (Radiation Sensor Systems)
Mid Sweden University, Faculty of Science, Technology and Media, Department of Electronics Design. (Radiation Sensor Systems)
Show others and affiliations
2013 (English)In: Journal of Instrumentation, ISSN 1748-0221, E-ISSN 1748-0221, Vol. 8, no 4, p. Art. no. C04004-Article in journal (Refereed) Published
Abstract [en]

A newly constructed solid state silicon dose profile detector is characterized concerning its sensitive profile. The use of the MEDIPIX2 sensor system displays an excellent method to align an image of an X-ray slit to a sample under test. The scanning from front to reverse side of the detector, show a decrease in sensitivity of 20%, which indicates a minority charge carrier lifetime of 0.18 ms and a diffusion length of 460 μm. The influence of diced edges results in a volumetric efficiency of 59%, an active volume of 1.2 mm 2 of total 2.1 mm2.

Place, publisher, year, edition, pages
2013. Vol. 8, no 4, p. Art. no. C04004-
Keywords [en]
Computerized Tomography (CT) and Computed Radiography (CR); Dosimetry concepts and apparatus; Solid state detectors; X-ray detectors
National Category
Electrical Engineering, Electronic Engineering, Information Engineering
Identifiers
URN: urn:nbn:se:miun:diva-18741DOI: 10.1088/1748-0221/8/04/C04004ISI: 000317462400004Scopus ID: 2-s2.0-84877780840Local ID: STCOAI: oai:DiVA.org:miun-18741DiVA, id: diva2:615941
Conference
14th International Workshop on Radiation Imaging Detectors; 1-5 July 2012; Figueira da Foz, PORTUGAL
Available from: 2013-04-12 Created: 2013-04-12 Last updated: 2017-12-06Bibliographically approved
In thesis
1. Advanced X-ray Detectors for Industrial and Environmental Applications
Open this publication in new window or tab >>Advanced X-ray Detectors for Industrial and Environmental Applications
2016 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

The new generation of X-ray free electron laser sources arecapable of producing light beams with billion times higherpeak brilliance than that of the best conventional X-ray sources.This advancement motivates the scientific community to pushforward the detector technology to its limit, in order to de-sign photon detectors which can cope with the extreme fluxgenerated by the free electron laser sources. Sophisticated ex-periments like deciphering the atomic details of viruses, filmingchemical reactions or investigating the extreme states of matterrequire detectors with high frame rate, good spatial resolution,high dynamic range and large active sensor area. The PERCI-VAL monolithic active pixel sensor is being developed by aninternational group of scientists in collaboration to meet theaforementioned detector requirements within the energy rangeof 250 eV to 1 keV, with a quantum efficiency above 90%.In this doctoral researchwork, Monte Carlo algorithm basedGeant4 and finite element method based Synopsys SentaurusTCADtoolkits have been used to simulate, respectively, theX-rayenergy deposition and the charge sharing in PERCIVAL. Energydeposition per pixel and charge sharing between adjacent pixelsat different energies have been investigated and presented.Novel methods for industrial and environmental applica-tions of some commercially available X-ray detectors have beendemonstrated. Quality inspection of paperboards by resolv-ing the layer thicknesses and by investigating orientation ofthe cellulose fibres have been performed using spectroscopicand phase-contrast X-ray imaging. It was found that, usingphase-contrast imaging it is possible to set burn-out like qualityindex on paperboards non-destructively. X-ray fluoroscopicmeasurements have been conducted in order to detect Cr inwater. This method can be used to detect Cr and other toxicelements in leachate in landfills and other waste dumping sites.

Abstract [sv]

Acceleratorbaserade röntgenkällor utvecklas ständigt, dessakan producera röntgenstrålning med miljarder gånger så högeffekttäthet som de starkaste konventionella röntgenkällorna.Därför finns en vetenskaplig utmaning att utveckla röntgende-tektorer som inte förstörs i de extrema flöden som genereras avdessa röntgenkällor. De visioner som finns för de nya källornaär t.ex.; att avbilda detaljer av virus ner på atomnivå, att filmakemiska reaktioner eller att undersöka extrema tillstånd hos ma-teria. Dessa typer av experiment kräver röntgendetektorer medhög bildhastighet, hög spatial upplösning och stort intensitets-omfång och stor aktiv sensoryta. Detektorsystemet PERCIVALsom bygger på aktiva pixlar med energiupplösning utvecklasinom ett internationellt vetenskapligt samarbetsprojekt. Må-let är att uppfylla detektorspecifikationerna för de nämndaexperimenten inom energiområdet 250 eV till 1 keV, med enkvantverkningsgrad över 90 %.I föreliggande vetenskapliga avhandlingsarbete har simule-ringar av energideponering i PERCIVAL-detektorn genomförtsbaserat på Monte Carlo-algoritmer och simuleringar av ladd-ningsdelning mellan pixlar har simulerats med hjälp av finitaelementmetoden. Därmed har energideponeringen per pixeloch laddningsdelningen mellan närliggande pixlar vid olikaenergier kunnat utredas och presenteras.I avhandlingen demonstreras nya lovande metoder för in-dustriella applikationer och miljöövervakning,därkommersiellttillgängliga röntgendetektorer kan användas. Kvalitetsövervak-ning av kartongtillverkning genom att mäta bestrykningstjock-lek och fiberorientering kan realiseras med energiupplöstaröntgenbilder eller faskontrastbilder i röntgenområdet. Det kankonstateras att med icke-förstörande provning, genom faskon-trastbilder, kan kvalitetsindexvärlden erhållas på samma sättsom kvalitetsindex kan erhållas från ”burn-out”-mätningar.Spektroskopiska mätningar av röntgenflourescens har genom-förts för att detektera krom (Cr) i vatten. Metodik för att detek-tera krom och andra giftiga metaller i lakvatten från deponioch annan lagring för giftigt avfall har utarbetats.

Place, publisher, year, edition, pages
Sundsvall: Mid Sweden University, 2016. p. 159
Series
Mid Sweden University doctoral thesis, ISSN 1652-893X ; 253
National Category
Electrical Engineering, Electronic Engineering, Information Engineering
Identifiers
urn:nbn:se:miun:diva-29264 (URN)STC (Local ID)978-91-88025-84-5 (ISBN)STC (Archive number)STC (OAI)
Public defence
2016-12-01, M108, Sundsvall, 10:15 (English)
Opponent
Supervisors
Available from: 2016-11-10 Created: 2016-11-10 Last updated: 2017-06-30Bibliographically approved

Open Access in DiVA

No full text in DiVA

Other links

Publisher's full textScopus

Authority records BETA

Thungström, GöranNorlin, BörjeReza, SalimKrapohl, David

Search in DiVA

By author/editor
Thungström, GöranNorlin, BörjeReza, SalimKrapohl, David
By organisation
Department of Electronics Design
In the same journal
Journal of Instrumentation
Electrical Engineering, Electronic Engineering, Information Engineering

Search outside of DiVA

GoogleGoogle Scholar

doi
urn-nbn

Altmetric score

doi
urn-nbn
Total: 578 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf