Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditionsVisa övriga samt affilieringar
2014 (Engelska)Ingår i: Journal of Instrumentation, ISSN 1748-0221, E-ISSN 1748-0221, Vol. 9, nr 4, s. Art. no. C04028-Artikel i tidskrift (Refereegranskat) Published
Abstract [en]
For clinical X-ray imaging, the detector performance under high flux conditions is very important, with typical flux rates for modern CT systems reaching 109 photons s-1 mm-2 in the direct beam. In addition, for spectral imaging a good energy resolution under these conditions is needed. This poses difficulties, since pulse pileup in the pixel electronics does not only affect the count rate, leading to a deviation from the otherwise linear behavior, but also degrades the spectral response of the detector, making k-edge subtraction and other contrast enhancement techniques less efficient. In this paper, we investigate the count rate capabilities and the energy response of the Medipix3RX chip under high flux conditions using 10 keV monochromatic photons. © CERN 2014.
Ort, förlag, år, upplaga, sidor
2014. Vol. 9, nr 4, s. Art. no. C04028-
Nyckelord [en]
Hybrid detectors, X-ray detectors
Nationell ämneskategori
Elektroteknik och elektronik
Identifikatorer
URN: urn:nbn:se:miun:diva-22044DOI: 10.1088/1748-0221/9/04/C04028ISI: 000336123800028Scopus ID: 2-s2.0-84899579747Lokalt ID: STCOAI: oai:DiVA.org:miun-22044DiVA, id: diva2:720654
Anmärkning
Language of Original Document: English
2014-06-022014-05-302017-12-05Bibliografiskt granskad
Ingår i avhandling